GE PROPRIETARY TO GE
D
IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
Chapter 4 Functional Checks 4-3
4-3-1-1 Boot Up (cont’d)
When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD,
Peripherals and the TMST.
4-3-1-2 Turn on the system
Press the Power On/Off switch at the front of the system once.
When the Power On/Off switch on the Control Panel is pressed once, the TMST starts and the software
code is distributed to initiate the scanner.
No status messages are displayed during this process.
4-3-2 Power Off/ Shutdown
NOTE: After turning off a system, wait for at least twenty seconds before turning it on again. The system may
not be able to boot if power is recycled too quickly.
4-3-2-1 TMST Power Down
To power down the system:
1.) Press the Power On/Off switch at the front of the system once.
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select key, select Shutdown.
Figure 4-2 Power On/Off Switch
Figure 4-3 System Exit Window
Power On/Off switch