GE PROPRIETARY TO GE
D
IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
Chapter 8 Replacement Procedures 8-55
Section 8-5 USB Cable Set (FRU P/N: 5315039)
8-5-0-1 Tools
• Common Phillips screwdrivers
8-5-0-2 Needed Manpower
• 1 persons, 15 minutes per cable
8-5-0-3 Preparations
• Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-5-0-4 Removal Procedure
1. Remove the covers and parts as applicable. Refer to the applicable section of this document.
2. Unplug the USB cable from both the ends and remove the cable from the machine.
8-5-0-5 Functional Checkout Procedure
See
Section
Functional Test Debrief Script
4-3-1 Power On/Boot Up
Proprietary Service Manual Direction 5308917-100, Section 8-5. Equipment passes all
required tests and is ready for use.
4-3-2 Power Off/ Shutdown