GE PROPRIETARY TO GE
D
IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
2-6 Section 2-2 - General Console Requirements
2-2-4 Scan Probe Environmental Requirements
Operation:10° to 40° C
Storage: 10° to 60° C
Storage: 10° to 70° C (LOGIQ P3 BT10)
NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32).
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
NOTICE
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-20 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
CAUTION
Have two people available to deliver and unpack the LOGIQ P3/LOGIQ P3 BT10.
Attempts to move the unit considerable distances or on an incline by one person could result
in injury or damage or both.