IRECTION 5308917-100, REVISION 8 LOGIQ P3 SERVICE MANUAL
8-24 Section 8-2 - Disassembly/Re-assembly of LOGIQ P3
8-2-11 Alpha Numeric Keyboard with cables (FRU P/N: 5315106-2)
8-2-11-1 Tools
• Common phillips screwdrivers
8-2-11-2 Needed Manpower
• 1 person, 30 minutes
8-2-11-3 Preparations
• Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-2-11-4 Removal procedure
1) Remove Keyboard Assembly.Refer section 8-2-7 on page 10
2) Remove the Alphanumeric Keyboard by unfastening the 10 screws (marked in Red color in
Figure 8-28 on page 8-24 ) holding it & the Alphanumeric keyboard cable.
Figure 8-28 Alpha Numeric keyboard PCB
8-2-11-5 Mounting procedure
1.) Install the new parts in the reverse order of removal.
A/N Keyboard
10 Screws