EasyManuals Logo

GE Voluson P8 Service Manual

GE Voluson P8
553 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #125 background imageLoading...
Page #125 background image
GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
3-60 Section 3-8 - System Configuration
3-8-3 On-Board Optional Peripherals
Mains outlet: AC outlet for on-board peripherals
Output voltage : 100~250V.
CAUTION
!! CAUTION:
Modification of voltage setting only by an authorized service person!
The maximum power consumption of equipment (inclusive of the color LCD monitor) connected
to these outlets must not exceed 400VA!
Table 3-8 Approved Peripherals
Device Manufacturer Model Connection Comment
Digital B/W printer SONY UP-D897 USB-Port
Digital Color Printer SONY UP-D25MD USB-Port
Digital Deskjet Printer
(Bluetooth)
HP Officejet H70 USB-Port (Bluetooth)
DVR GEUK Dedicated Connector
ECG Module Norav Marlin/ECGUSB1D USB-Port
Footswitch Whanam Elec. USB-Port

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based