GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 4 - Functional Checks 4-21
4-4-5-2 Functions after the Acquisition
Figure 4-14 Sectional Planes and Image Rendering
Table 4-8 Functions after the 3D Acquisition
Step Task Expected Results
1
PW MODE
rotary control: Rotation about X-axis and Y-axis of the reference
image.
M MODE
rotary control: Rotation about the Z-axis of the reference image.
2
CF MODE
rotary control: CF-Gain and CF-Mode.
TRACKBALL
: Movement along X- and Y-axis of the reference image.
3
NICHE
Parts of the orthogonal sections A, B and C are compiled to a 3D-section aspect.
The aspect shows a quasi spatial cut into the reference image.
4
REF. IMAGE To select the Reference image among A, B, C or 3D.
5
STEER/3D INIT
In Color flow, ROI can be steered using this button (in Linear probes)
In 3D/4D Mode, the button resets the image to its initial position.