7-34 Section 7-9 - Hardware Tests
7-9-1 Audio(System) Interface Test.
Test of the Audio path from system to Speaker and AudioOutput can be performed.
1.) Select the AUDIO(SYSTEM) INTERFACE TEST button in the Diagnostics Tool window.
2.) Check if there is hearable audio sound from Speaker and AudioOutput port on rear panel.
7-9-2 Audio(Doppler) Interface Test.
Test of the Audio path from Doppler Audio generator to Speaker and AudioOutput can be performed.
1.) Select the AUDIO(DOPPLER) INTERFACE TEST button in the Diagnostics Tool window.
2.) Check if there is hearable beep audio sound from Speaker and AudioOutput port on rear panel.