GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-53
7-12-2 Noise disturbs the Image
Figure 7-54 Noise disturbs the Image - Troubleshooting
Noise is shown in image.
YES
NO
Is there any electrical
device that affect the
Scanner?
Keep the scanner away from those
electrical devices.
use another wallsocket
Check probes.