GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
8-20 Section 8-8 - Replacement or Activation of Options
8-8-1 Operation for activating Options
1.) Press the F2 key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION
and then click the OPTIONS tab (see:
Figure 8-9 on page 8-11.)
8-8-1-1 Operation for installing a “Demo Key” or a “Permanent Key”:
1.) Position the cursor inside the input field desired and press the
right/left trackball key.
2.) If one exists, clear/edit the current key code.
3.) Enter the encrypted serial code with the keyboard and click on SUBMIT
. (Code will be checked.)
4.) Click the SAVE&EXIT
button.
NOTE: After activating a key code, restart (turn off and on) the Voluson® P8 / Voluson® P6 system.

Table of Contents

Other manuals for GE Voluson P8

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals