4-34 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
Section 4-5
Backup and Restore Database, Preset Configurations and Images
1.) Press the F2 key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP
and then click the SYSTEM CONFIGURATION tab.
The
SYSTEM CONFIGURATION page is subdivided in 2 main groups:
A.) Image Settings Only
Section 4-5-1 "Save Image Settings Only" on page 4-35
Section 4-5-2 "Load Image Settings Only" on page 4-36
B.) Full System Configuration
Section 4-5-3 "Save Full System Configuration (Full Backup)" on page 4-39
Section 4-5-4 "Load Full System Configuration (Full Backup)" on page 4-41
Section 4-5-5 "Delete Full System Configuration (Full Backup)" on page 4-43
The Image Settings and/or Full System Configuration can be saved to the following destinations:
D: partition of internal hard disk
DVD/CD+R/RW
Mapped Network Drive, see: Section 3-12-1 "Map Network Drive" on page 3-71
Any other drive connected to the system (e.g.; USB-Stick or external hard disk drive)
Note:
This function is only available in the Full Backup utility.
For further details review: Section 3-5-9 on page 3-21
CAUTION
!! CAUTION:
It is recommended to Backup the Full System Configuration (Section 4-5-3 on page 4-39) and the
Image Archive (Section 4-5-6 on page 4-44) once a week.
Figure 4-19 System Setup - Backup - SYSTEM CONFIGURATION page
Image Settings only
Full System
Configuration

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals