GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-25
7-5-7 CSD: Configuration
7-5-7-1 To configure Service Platform
1.) Ask the hospital's network Administrator for the Proxy Server Address and the Proxy Server Port.
2.) Select the Configuration page, then double-click “InSite ExC Agent Configuration”.
3.) Fill out all bold stated “mandatory” fields.
4.) Select PRODUCT server in the list.
5.) Enter Proxy Server Address and Proxy Server Port.
6.) Click the SUBMIT CHANGES
button and then close the service page.
NOTICE
!! NOTICE:
If a Proxy server is available, the information MUST be properly entered, otherwise remote control
does not work. There is no possibility to detect proxy server information automatically.
Figure 7-29 Common Serive Desktop - Configuration
Example shows
fictional data!
NOTE:
“Device” and “Serial Number
is KP6MENG025
(prefix KP6 or KP8 +
system´s serial number
without prefix-letter).
Select PRODUCT
server in the list
Enter proxy server
information if it requires

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals