GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
7-52 Section 7-12 - Troubleshooting Trees, Instructions and Tech Tips
7-12-1 System does not boot up
Figure 7-53 System does not Power On / Boot Up
System doesn’t boot
Is Standby Switch
illuminated?
Press Standby Switch till it
changes to green System
starts booting
Is Standby Switch
orange?
Is there any picture on
the monitor ?
Other error. Please contack
the service engineer
Is there any picture on
the monitor ?
- Power Cord is connected to Mains
- Power Switch is turned on
(on the backside of the system)
Verify that
Press Standby Switch for
approximately 5 sec.
untill system is down
Check cables between
Console and Monitor
No
No
No
No
Yes
Yes
Yes
Yes

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals