Chapter 5 - Components and Functions (Theory) 5-17
tissue scatters, image speckle characteristics such as brightness, density or size have no apparent
value in determining tissue structure or related properties. The elimination of or significant reduction in
speckle improves the quality or diagnostic potential of the image.
The method applied in the subject modification utilizes a nonlinear diffusion filtering technique that
permits effective speckle reduction in real time. The speckle reduction filter is available to the user in all
B-mode imaging, independent of the transducer used.
5-2-4-6 SonoNT
SonoNT is an additional function for manual NT (Nuchal Translucency) measurement. This function
supports the user to find the correct position for the NT measurement. The user can switch between NT
Method “Manual” and “Sono NT” (semi-automatic).
A Box has to be placed for the NT-ROI. Then the NT-distance is calculated automatically, a graphic
(yellow head-image) and the NT-result are displayed.
If no result is found a temporary warning “No valid NT-distance found!” is displayed.

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals