GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
4-4 Section 4-3 - General Procedure
4-3-2 Power Off / Shutdown
4-3-2-1 Scanner Shutdown
1.) If not already in read mode, freeze the image.
2.) Press the ON/OFF button (see: Figure 4-2) on the control panel. Following dialog appears.
3.) Select the SHUTDOWN
button. The system performs an automatic full shutdown sequence.
NOTE: Full shut down is also performed when pressing the ON/OFF button on the control panel twice.
4.) Switch OFF the Circuit Breaker at the rear of the system.
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
ON/OFF button. So the auxiliary equipment need not to be switched ON/OFF separately.
5.) After complete power down, unscrew the 2 screws and remove the pull-out protection to disconnect
the main power cable from the system or unplug it from the AC wall outlet socket.
NOTICE
!! NOTICE:
After turning off a system, wait at least 10 seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
Figure 4-3 Shutdown, Restart or Cancel
WARNINGWARNING
!! WARNING:
Disconnection of the Main Power Cable is necessary!
For Example: When repairing the system.
Figure 4-4 Circuit Breaker, pull-out protection and main power cable at rear of system
Circuit breaker
Pull-out protection
After unscrew 2 screws

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals