GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
4-46 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-6-2 Load Image Archive
1.) Press the F2
key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP
and then click the IMAGE ARCHIVE tab.
4.) Click the LOAD
button (see: Figure 4-32 on page 4-44).
5.) Choose the Destination.
6.) Click on the backup to be restored (additional information is displayed in the table).
7.) Select the NEXT
button. The following window will be displayed.
8.) Select (check mark) the image archive data to be restored to the Voluson® P8 / Voluson® P6
system.
9.) Click the NEXT
button and then select YES to start, or NO to cancel the restore procedure.
Figure 4-35 Image Archive Load - choose destination
Figure 4-36 select image archive data to be restored
click [Next] button
choose destination
select backup file
click [Next] button
select image archive
data to be restored

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals