Page #280 background imageLoading...
Page #280 background image
GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-73
5-15-6-3 SSHD
By selecting this item, a SSH-daemon is started on the system.
The [SSHD] function is reserved for extended troubleshooting. Not intended for use in the field.
5-15-6-4 Restore
This function takes the image stored on the Rescue (R:) partition to replace the existing image of the
“C” partition.
NOTE: Data on the D: [User] partition (e.g., Patient Archive images, Measurement results, Patient Data and
Reports) will NOT be modified!
5-15-6-5 Initialize Hard
This function performs operations which result in complete data loss of all files on C:\ and D:\
partition. The image on the on the Rescue (R:) partition of the hard disk is used for restoring C:\.
5-15-6-6 ServiceTools
Service Tools Menu, see: Section 5-15-7 on page 5-76
5-15-6-7 Exit
Return to the “Advanced Functions Menu”.
5-15-6-8 Shell
By selecting this item, a Bash-shell is opened. Enter “exit” to return to the main menu.
The [Shell] function is not intended for use in the field.
NOTICE
The [Restore] function should only be performed if NO System DVD is available!
If you have a System DVD, please select the [Restore] function that is available in the
CD-ROM Functions Menu; see:Section 5-15-5 on page 5-61
NOTICE
The [Restore] function should only be performed if NO System DVD is available!
If you have a System DVD, please select the [Restore] function that is available in the CD-ROM
Functions Menu; see:Section 5-15-5 on page 5-61), which is only available when booting from the
System DVD
WARNINGWARNING
All data on D:\ (User) partition (e.g., Patient Archive images, Measurement results,
Patient Data and Reports) will be deleted and re-initialized.
NOTICE
The [Initialize] function should only be performed if NO System DVD is available!
If you have a System DVD, it is highly recommended to use the [New Disk] function (see:Section 5-
15-5-4 on page 5-64), which is only available when booting from the System DVD

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals