Page #371 background imageLoading...
Page #371 background image
GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
8-8 Section 8-3 - System Software - Installation/Upgrade Procedure
Section 8-3
System Software - Installation/Upgrade Procedure
8-3-1 Introduction
8-3-2 Manpower
One person ~ 1 hour (depends on contents of System DVD, peripherals, etc.)
8-3-3 Tools
System DVD
8-3-4 Preparations
Before performing the Software Upgrade:
make sure that all system functions are working correct
check the current Application Software version (Figure 8-9) and the installed Options (Figure 8-9)
if the currently installed software has to be upgraded by a newer version, and the system DVD being
used is SW 13.0.0 or higher, calculate new software specific “Permanent key” in OKOS.
Please contact your local distributor or GE service representative to get the necessary key.
NOTE: It is NOT necessary to save Full System Configuration (Full Backup) prior to the upgrade.
All existing User Programs, 3D/4D Programs and Auto Text settings remain untouched!
NOTICE
!! NOTICE:
To update the Systems C:\ image via the FMI FROM DVD button in the System Setup SERVICE page,
see: Section 8-3-5 "System Software - Installation Procedure (FMI from DVD)" on page 8-12.
The Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS Patches, etc.) depend on the contents of the System DVD being used.

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals