GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 4 - Functional Checks 4-37
4-5-2-2 Load “Complete Backup”
NOTE: Following procedure should be used, if the user prefers factory default settings, which are adapted for
the installed Application Software version.
1.) Perform Section 4-5-2-1 "Preparations" on page 4-36.
2.) Select the “Complete Backup” (marked blue; see: Figure 4-22 below) and click the >>
button to
copy the Complete Backup into the Load Data field.
3.) Click the LOAD
button to start the loading procedure of the complete backup into the system.
The US Application Software restarts.
Figure 4-22 Load Backup Data
>> button
[Load] button
Complete Backup