GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 4-16 Section 4-3 - General Procedure
4-3-11 Backup and Restore Database, Preset Configurations and Images
NOTE: When upgrading from R1.0.x to R2.0.x or R3.0.x , preset cannot be transferred. Refer to the
table below.
4-3-11-1 Formatting Media
1.) To format the backup media, CD-R or DVD-R, select the UTILITY
key on the Front Panel.
2.) Select CONNECTIVITY
, then RERNOVABLE MEDIA. Properly label and Insert the backup
media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-13. It is best to use all capital letters with no
spaces or punctuation marks.
Table 4-4 LP6/LP6pro BT11(R3.x.x) compatibility
Preset
Compatibility
R1.x.x R2.x.x R3.x.x
Patient Archive
Yes Yes Yes
Report Archive
Yes Yes Yes
Imaging Presets
No No No
Connectivity Configuration
No No No
Measurement Configuration
No No No
Comment/Body Pattern Libraries
No No No
Report Templates (Same Software Version Only)
No No No
3D/4D
No No No
Key Macro
No No No
All Others
No No No