8-2-2-3 17" LCDMON OSD Key Assy Replace Procedure (cont’d)
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-2-4 Mounting Procedure
Install the new parts in the reverse order of removal
Table 8-3 Functional Tests
Service Manual
Section Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-2. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-6-2
17" OSD Key Assy Function Validation
Procedure
Section 10-5-5
Physical Inspection