GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 8-44 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-11 TGC key assy / TGC Knob Set
8-2-11-1 Tools
Common Phillips screwdrivers
8-2-11-2 Preparations
Shut down the system and switch off the main breaker.
8-2-11-3 Removal procedure
1.) Remove the CFUNCKEY Assy. Refer to the 8-2-6 "Main Keyboard Assy" on page 8-37.
2.) Remove the TGC Knobs.
3.) Unscrew 4 screws (1-4).
Figure 8-81 8 TGC Knobs
Figure 8-82 TGC PCB Assy in Keyboard
12
34