GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 7-12 Section 7-1 - Overview
7-5-3 Capturing a Screen (cont’d)
5.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the
image on the hard disk or CD-R or USB.
The InSite connection will have access to the export folder on the “D” drive to retrieve these images.
7-5-4 Reset the P2 Key to Customer’s Functionality
If the customer had programmed the P2 Key to a function other than screen capture, restore that
functionality recorded in section 7-5-1 on page 9. Refer to the Figure 7-7.
1.) Select Utility from the Keyboard.
2.) Select Connectivity from the Menu.
3.) Select the Buttons tab on the Connectivity screen.
4.) In the Button field, select Print2.
5.) In the “Physical Print Buttons”box, select the options recorded in step 5, Section 7-5-1.:
*Format
* Image Frames
* Capture Area
* Compression
Figure 7-10 Save Dialog Box