GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 4-48 Section 4-9 - Board Function Checks
4-9-9 BP128 Assy Function Validation Procedure
1.) Turn on the system
2.) All of attached ASSY should be tested.
3.) Refer to the 4-9-5 "SYSCONML Assy Function Validation Procedure" on page 4-45.
4.) Refer to the 4-9-7 "SOM Function Validation Procedure" on page 4-47.
5.) Refer to the 4-9-2 "DTRX64II Assy Function Validation Procedure" on page 4-39.
6.) Refer to the 4-9-1 "PI128II Assy function check procedure" on page 4-37.
7.) Refer to the 4-9-6 "APS II Assy PCB Function Validation Procedure" on page 4-46.
8.) Refer to the 4-9-4 "CWD64 Assy Function Validation Procedure" on page 4-42.(Optional)