LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-101
8-2-47 Read and Write Serial Info
8-2-47-1 How to start serial no read/write program
1.) Plug the service dongle on the usb port on rear panel.
2.) Turn on the system.
3.) When the Start Application dialog box appears, press [F2] before it automatically starts
Echoloader.)
NOTE: If maintenance access box does not appear and the scanning mode starts, just insert the
Service Dongle and press “CTRL+ALT+R”.
4.) When the Maintenance Access dialog box pops up, type in the service password.
Figure 8-147 Start Application Dialog Box
Figure 8-148 Maintenance Access Dialog Box