GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 5-38 Section 5-1 - Overview
5-8-3 Sending RFS Procedure
Service Platform should be turned off in order to open RFS window.
System must be connected to network, configured and checked out to be able to send the RFS.
1.) Right Click on “Connect to GE” icon.
All the fields that are marked with asterisk * are mandatory and should be filled in order to send
the RFS request to On Line Center.
2.) Fill out the Last name, First name, and Phone number. Refer to the Figure 5-40.
3.) Extension, Email, and Other system ID fields are optional.
4.) Select Problem Type.
If Problem is about “No Boot”, “No Image”, “Error message”, “Lock up”, “Probe”, “Peripherals”, select
Service” for the Problem Type.
If Problem is about “Presets”, “Reports”, “Measurements”, or “Probe not recognized”, select
Applications” for the Problem Type.
5.) Select the Problem Area.
6.) Write the detailed problem description in the Problem Description field.
Figure 5-39 Contact GE Icon
Figure 5-40 RFS window

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 Pro Series and is the answer not in the manual?

GE LOGIQ P6 Pro Series Specifications

General IconGeneral
BrandGE
ModelLOGIQ P6 Pro Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals