GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 8-118 Section 8-4 - Software Installation
8-4-2-2 Driver Checks
1.) On the window XP desktop, left click on My Computer then right click on Properties.
2.) The System Properties window appears. Select Hardware > Device Manager.
3.) Verify that ?, X, or ! is not displayed.
If X or ! is displayed:
- Left-click on the device with X mark, then right -click on Enable
- Left-click on the device with ! mark, then right click on Properties to perform the device driver
recognition. When some files are required, designate the file in the director
ies of
C:\WINNT\SYS
TEM32 or C:\WINNT\SYSTEM32\DRIVERS. If they are not found, search
them using a window search function
Figure 8-176 Device Manager