Oscilloscope Reference Delayed trigger system
Delayed trigger system
You can trigger with the A (Main) trigger system alone or you can combine the A (Main) trigger with
the B (Delayed) trigger to trigger on sequential events. When using sequential triggering, the A trigger
event arms th
e trigger system, and the B trigger event triggers the instrument w hen the B trigger conditions
are met.
AandBtrigg
ers can (and typically do) have separate sources. The B trigger condition can be based on a
time delay or a specied number of counted events.
For more de
tails on delayed triggering, clickhere
(see page 909).
What do you want to do next?
Learn about advanced triggering. (see page 702)
Advanced triggering
You can check the advanced trigger status in the readout. The r eadout indicates the trigger type and then
shows sources, levels, or any othe r parameters that are important for the particular trigger type.
702 DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series

Table of Contents

Other manuals for Tektronix DPO70000C/DX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000C/DX Series and is the answer not in the manual?

Tektronix DPO70000C/DX Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000C/DX Series
CategoryTest Equipment
LanguageEnglish

Related product manuals