Oscilloscope Reference Gaussian window
What do you want to do next?
Learn about using spectral analyzer windows. (see page 771)
Gaussian window
This is the default window function (see the next gure). It is unique in that the time-domain shape o f an
exponential Gaussian function transforms into a Gaussian exponential shape in the frequency domain.
NOTE. This window provides optimal localization in both the time and the frequency domain
What do you want to do next?
Lea
rn about using spectral analyzer windows.
(see page 771)
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 775