About Tektronix oscilloscopes TekScope Recovery Report Utility
Data contained in the report
When the utility is finished running, if these files exist on the instrument, the report will contain copies
of these specific files:
C:\Windows\Sysnative\winevt\ logs\Application .evtx
C:\programdata\tektronix\ISD .XML
C:\programdata\Tektronix\Tek Scope\Calibratio n\console.log
C:\program
data\Tektronix\TekScope\Cal ibration\console .logOld
C:\programdata\Tektronix\Tek Scope\Calibratio n\calSPCConst.d at
C:\programdata\Tektronix\Tek Scope\Calibratio n\caldiag.log
C:\programdata\Tektronix\Tek Scope\Calibratio n\caldiag.logOl d
C:\programdata\Tektronix\Tek Scope\Calibratio n\error.log
When the utility is finished running, the zipped report file will also contain copies of all files found in the
following directories:
C:\programdata\Tektronix\Tek Scope\Internal\* .*
C:\programdata\Tektronix\Tek Scope\Calibratio n\ CalDramDump\*.*
C:\programdata\Tektronix\Tek Scope\Calibratio n\ CalDramDumpFact\*.*
Data deleted by the utility
Runni
ng this utility will delete the following files:
C:\users\<current_user>\AppD ata\LocalLow\Tek tronix\ tekScope\Internal\*.*
Because TekScope was having difficulties, some files may have been corrupted. Deleting these files will
allow TekScope to regenerate them on the next start-up. Deleting these files may resolve the initial problem.
The utility log file
Yo
ucanseealogfile of the operation performed by the recovery report utility at:
C:\Temp\TekScopeRecoveryRepo rtUtility_log.tx t