GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A233 - Lift Screw Assembly Top Brake
Page no. 1279 Chapter 9
JC-DR-A-233.fm
Job Card D/R A233 - Lift Screw Assembly Top Brake Chapter 9
1SUPPLIES
Lift Screw Assembly Top Brake
2 APPLICABILITY
This Job Card is only applicable to systems with a Type 2 Lift Screw Top Assembly. A Type 2 Lift Screw
Top Assembly has a silver L-bracket withthreaded holes, as indicated below.
3TOOLS
Standard Tool Box
4 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 1 hour
5 SAFETY PRECAUTIONS
WARNING
The Gantry must be powered OFF.
6 PREREQUISITES
Switch OFF the Gantry power.
Type 1
Type 2
Type 3