EasyManuals Logo
Home>GE>Medical Equipment>Senographe DS

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1337 background imageLoading...
Page #1337 background image
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job card D/R A252 - Compression Trolley Kit
Page no. 1337 Chapter 9
JC-DR-A-252.fm
Job card D/R A252 - Compression Trolley Kit Chapter 9
1SUPPLIES
Compression Trolley Kit (comprising Paddle Holder, and W304B Cable)
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4 SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5 PREREQUISITES
• Switch off the Gantry power.
6 PROCEDURE
1. Remove the Compression Carriage covers (refer to section 8-1 in Job Card PHY A044 - Remove/
Reinstall Gantry Covers on page 523).
2. Remove the Compression trolley (refer to section 8-2 in Job Card PHY A044 - Remove/Reinstall
Gantry Covers on page 523).
3. Change the old Compression trolley for a new Compression trolley (refer to section 8-2 in Job Card
PHY A044 - Remove/Reinstall Gantry Covers on page 523).
4. Replace the original Compression Carriage covers (refer to section 8-1 in Job Card PHY A044 -
Remove/Reinstall Gantry Covers on page 523).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals