GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A314 - External CD-RW Unit
Page no. 1439 Chapter 9
JC-DR-A-314.fm
Job Card D/R A314 - External CD-RW Unit Chapter 9
1 APPLICABILITY
This Job Card is not required if the internal CD-R writer is present.
Applicability code B. See Compatibility codes on page 68.
2SUPPLIES
CD-RW unit
3TOOLS
Standard Tool Kit
4 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
5 SAFETY PRECAUTIONS
None
6 PREREQUISITES
The installed AWS software must be ADS version 29.0 or later.
7 PROCEDURE
7-1 Configure the CD-RW unit
Switches on the rear panel of the new CD-RW unit must be correctly set before installation. They are not
accessible after the Control Station covers have been installed.
• Check that the power switch at the rear of the new
CD-RW unit is set to to ON (I). A sticker may have
been placed over the switch to prevent it from being
set off.
• Check that the SCSI ID switch is set to 2. If neces-
sary, change the setting up or down by inserting a
small flat screwdriver or similar tool into one of the
two slots above and below the switch.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals