Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A429 - Conditioner Air Filter Cleaning
Page no. 1807 Chapter 11
JC-PM-A-429.fm
Job Card PM A429 - Conditioner Air Filter Cleaning Chapter 11
1SUPPLIES
Roll of adhesive tape
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 5 minutes
4 SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment and switch OFF the Senographe hospi-
tal supply circuit breaker. Lock it in the OFF position or mark it clearly with a warning sign
to prevent others from applying power.
5 PREREQUISITES
• Shut down and switch off the equipment as per the Warning above.
• Remove the Generator Cabinet covers as required for access to the Conditioner. Refer to Job Card
PHY A042 - Remove/Reinstall Generator Covers on page 513.
6 PROCEDURE
To clean the Conditioner air filter, proceed as follows:
1. Loosen, without removing, screw (1) securing air
duct (2) in place, then carefully remove air duct.
2
1

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals