GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Page no. 729 Chapter 7
JC-ELE-A-043.fm
Job Card ELE A043 - AWS Monitor Calibration Chapter 7
1SUPPLIES
None
2TOOLS
A calibrated Luminance Meter that permits measuring the luminance of the monitor in contact (Direct)
mode as illustrated below.
• Siemens Serial Spot Meter (P/N 2319223) must be used with the foam ring attached, and must be
configured in Direct mode.
• The Away/Direct button toggles between away and direct modes. To select the Direct mode, keep
pressing the Away/Direct button until D is displayed on the lower left corner of the Spot Meter
screen.
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
Remove canon and place foam ring
Luminance Meter
Foam Ring
Direct mode selected
Away/Direct mode
button selector