Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
Page no. 1503 Chapter 9
JC-DR-A-321_DS.fm
Job Card D/R A321 - IDC Computer Chapter 9
1SUPPLIES
• IDC Computer - Dione 4 or Dione 5, depending on your system level as follows:
- Go to the Service Desktop home page (see Launching the Service Desktop on page 120).
- Consult Table 1 on page 63 in the section Component FW/SW to ADS Version compatibility
Matrix on page 63. Determine the system level from the ADS/IDC/Generator CPU versions dis-
played on the Service Desktop home page, and by consulting Table 1 on page 63. The system
level will be defined by one of the System Level (Model Number) columns in Table 1 on page 63.
- If your system is at pre-Penduick level, order the Dione 4 (2375961)
- If your system is at Penduick or above level, order the Dione 5 (5324091)
• Floppy disk or CD-ROM with parameters for the currently installed image receptor
2TOOLS
Standard Tool Kit
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 90 minutes
4 SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment correctly and switch OFF the
Senographe system hospital supply circuit breaker. Lock it in the OFF position and mark it
clearly with a warning sign to prevent others from applying power.
5 PREREQUISITES
• Remove Control Station omega cover. See Job Card PHY A040 - Remove/Reinstall Control Station
Covers on page 503.
• Backup the system configuration to a floppy diskette or a CD-R. See Job Card ELE A052 - Back-up/
Restore Procedure on page 771. If a current backup is not available, it will be necessary to carry out
all detector calibration procedures after the replacement.