EasyManuals Logo
Home>GE>Medical Equipment>Senographe DS

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1357 background imageLoading...
Page #1357 background image
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A264 - Tilt Sensor
Page no. 1357 Chapter 9
JC-DR-A-264.fm
Job Card D/R A264 - Tilt Sensor Chapter 9
1SUPPLIES
Zero Tilt Sensor and Cable W405
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 2 hours
4 SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5 PREREQUISITES
• Rotate the arm to 90° for access to the Tilt Sensor (1).
• Switch off the Gantry power.
• Remove the tube cover (refer to Job Card PHY A044 -
Remove/Reinstall Gantry Covers on page 523).
1

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals