GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A245 - Manual Compression Belt
Page no. 1299 Chapter 9
JC-DR-A-245.fm
Job Card D/R A245 - Manual Compression Belt Chapter 9
1SUPPLIES
Manual Compression Belt
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4 SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5 PREREQUISITES
• Lower the arm to obtain good access to
the Manual Compression Belt (1).
• Switch off the Gantry power.
• Remove the compression covers (Refer to
Job Card PHY A044 - Remove/Reinstall
Gantry Covers on page 523).