GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A425 - IDC Air Filter Cleaning
Page no. 1803 Chapter 11
JC-PM-A-425.fm
Job Card PM A425 - IDC Air Filter Cleaning Chapter 11
1SUPPLIES
None
2TOOLS
• Vacuum Cleaner
• Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 10 minutes
4 SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment and switch OFF the Senographe hospi-
tal supply circuit breaker. Lock it in the OFF position or mark it clearly with a warning sign
to prevent others from applying power.
5 PREREQUISITES
• Shut down and switch off the equipment as per the Warning above.