GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A208 - Gantry Curtains
Page no. 1193 Chapter 9
JC-DR-A-208.fm
Job Card D/R A208 - Gantry Curtains Chapter 9
1SUPPLIES
Kit containing upper and lower Gantry Curtains
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 60 minutes
4 SAFETY PRECAUTIONS
WARNING
While performing this procedure, you must:
Power off the Gantry.
Wear safety glasses and gloves.
5 PREREQUISITES
• Switch off the Gantry power.
• Remove the front and column covers (refer to Job Card PHY A044 -
Remove/Reinstall Gantry Covers on page 523.
• Switch on the Gantry power.
• Use the lift command buttons to move the arm vertically until the two
springs (1) are fully visible in the opening at the rear of the column
frame.
• Switch off the Gantry power.