Job Card PHY A044 - Remove/Reinstall Gantry Covers
Page no. 523 Chapter 6
JC-PHY-A-044.fm
Job Card PHY A044 - Remove/Reinstall Gantry Covers Chapter 6
1SUPPLIES
None
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 20 minutes to remove covers, and 30 minutes to reinstall covers
4 SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment and switch off the Generator hospital
supply circuit breaker. Lock it in the off position or mark it clearly with a warning sign to
prevent others from applying power.
Never lay a cover down with the painted side on a hard surface.
5 PREREQUISITES
None
6OVERVIEW
For an overview of the covers, refer to Gantry Covers on page 189.
List of covers:
• Column Covers (section 7)
- Cable Exit Cover
- Top Column Cover (Left and right)
- Bottom Column Covers (Rear and front)
- Front covers (Left and right)
• Compression and Arm Covers (section 8)
- Compression Carriage Cover
- Compression Arm Covers (Left and right)
• Tube Head Covers (Left and right) (section 9)
• Detector Covers (Upper and lower covers) (section 10)
There are different types of covers
• Pilot covers
• Series covers which are divided into two kinds:
- Flexible covers
-Rigid covers
The procedure is valid for all covers except where specified.