GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A042 - Bar Code Scanner (Option)
Page no. 715 Chapter 7
JC-ELE-A-042.fm
Job Card ELE A042 - Bar Code Scanner (Option) Chapter 7
1SUPPLIES
Bar Code Scanner
2TOOLS
None
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 17 minutes
4 SAFETY PRECAUTIONS
None
5 PRE-REQUISITES
The system must have been correctly installed and normal application software must be operational.
6 PROCEDURE
6-1 Introduction
The bar code scanner is a fast and accurate tool for scanning bar code symbols in AWS applications. It
allows patient information to be entered rapidly and accurately.
Available with an 80 mm CCD scanning window and an integrated keyboard decoder, the scanner cap-
tures an entire bar code symbol instantly. The integrated decoder then transmits the data to the AWS
workstation as if the data was entered from the AWS keyboard.
Supported bar code types are:
- Code 3 of 9 (standard & full-ASCII)
- EAN 13
- Code 128

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals