GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A007 - IDC Error Messages
Page no. 937 Chapter 9
ERR-SUB-A-007.fm
ERR SUB A007 - IDC Error Messages Chapter 9
1 HOW TO VIEW THE ERROR LOG
Errors reported by the IDC diagnostic routines are recorded in a common system error log. The common
system error log also contains error messages generated by the AWS and Gantry/Generator applica-
tions.
To view the error log content, right click the AWS Screen background; select Service Tools/Service
Desktop/Error Log from the drop-down menu.
The most recent error code generated by the IDC can also be viewed on the display on the rear of the
IDC computer.
2 IDC DIAGNOSTICS
Each time that the IDC sub-system is booted (powered up or reset), it carries out a series of diagnostic
self-tests, first under the control of the BIOS, then under software control. As each test is performed, a
corresponding code appears on the four-digit green LCD panel at the rear of the IDC Computer.
2-1 BIOS Diagnostic Display
While the BIOS tests are running, the LCD displays two blank digits followed a two-digit code corre-
sponding to the current test. If a test detects an error or fails to complete, the tests are halted and the
code of the failing test remains displayed.
2-2 Software Diagnostic Display
• Three sets of software diagnostics are carried out. While these tests are running, the display shows
X000, where X may be 1, 2, or 3.
• If all tests are completed without failure, the display shows FFFF.
• If a test fails, the display shows a code in the form XXFF for two seconds. XX is a code correspond-
ing to the failing test. The tests then continue.
• On completion of the tests after a failure occurs, the display shows F0FF.
3 ERROR DETECTION
• If the IDC halts with a display in the format FFFF, all power-up tests have passed without failure.
• If the IDC halts with a display in the format --XX, where XX can be any number, a BIOS test has
failed. If the problem occurs again after reset, the recommended action is to change the IDC Com-
puter.
• If the IDC halts with a display in the format F0FF, a software test has failed. Reset the IDC and watch
the LCD display.
If the failure is repeated, a code XXFF appears for two seconds. Note the error code XX.
Table 1 gives a list of IDC diagnostic error codes, with recommended sequences of actions.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals