GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Product history and compatibility
Page no. 63 Chapter 3
Product-historyDS.fm
4 COMPONENT FW/SW TO ADS VERSION COMPATIBILITY MATRIX
The components in the Senographe system must have a certain version of firmware or software in order
to work correctly at a particular system level. The following table defines the compatibility matrix between
the software/firmware version required for each of these components for each system level.
TABLE 1 - COMPONENT FW/SW COMPATIBILITY MATRIX
System Level (Model Number)
Component IS4-3
(2383168-5
and lower*)
Orion1
Nephtys
M412s1
Nephtys M413
(2383168-6)
Penduick
(2383168-7)
Z400
(2383168-8)
ADS Hardware Computer SB150 or U20 U20 U20 Z400
Detector Type SFOV1 or
SFOV2
SFOV1 or
SFOV2
SFOV1 or
SFOV2
SFOV1 or
SFOV2
IDC Hardware Computer Dione 4 Dione 4 Dione 5 Dione 5
ADS Software Kit 53.10.1 (U20)
43.10.1 (SB150)
53.10.10 53.40 54.10
IDC Software Kit 4.2.6 ***** 4.2.6 ***** 10.0.5 10.0.5
Generator CPU Board
(DMR)
1.10 3.02 4.02 4.02
Gantry CPU Board
(Poseidon)
NSC.1.1.7 or
NSC.1.1.8 **
NSC.1.2.5 NSC.1.2.15 NSC.1.2.15
Interface Board FPGA 4A FPGA 4A FPGA 4A FPGA 4A
Rotation Board CPLD 5.1.2 5.1.2 5.1.2 5.1.2
DSP 5.13.5 5.13.5 5.13.5 5.13.5
HC12 5.9.4 5.9.4 5.12.4 5.12.4
Lift Board CPLD 5.0.1 or 10.0.1*** 10.0.1 10.0.1 10.0.1
DSP 5.6.3 5.6.3 5.6.3 5.6.3
HC12 5.10.2 5.15.2 5.16.2 5.16.2
Compression Board CPLD 5.1.4 5.1.4 5.1.4 5.1.4
HC12
5.7.1 5.7.1 5.12.1 5.12.1
Tilt Board CPLD 5.1.5 5.1.5 5.1.5 5.1.5
HC12 5.4.0 5.4.0 5.4.0 5.4.0
Collimator DSP Rev1026 Rev1026 Rev1026 Rev1026
PDU Board HC12 4.0.7 or 5.0.7 **** 5.0.7 5.0.7 5.0.7
Bucky Board CPLD 5.0.3 5.0.3 5.0.3 5.0.3
HC12
5.4.6 5.4.6 5.4.6 5.4.6
Gantry CD-ROM to use to get
required system level
5174994-7 5174994-9 5174994-11 5174994-11
*- With the complete rollout of FMI12075, all systems in the field are at least at an IS4-3 system level.
** - Systems that received FMI12075 have NSC.1.1.7 and IS4-3 Forward Production systems will have
NSC.1.1.8. You must use Gantry CD-ROM to ensure that the Gantry CPU Board is at NSC.1.1.8.
*** - Systems that received FMI12075 may have Lift CPLD 5.0.1 and IS4-3 Forward Production systems will
have Lift CPLD 10.0.1.
**** - Systems that received FMI12075 may have PDU HC12 4.0.7 and IS4-3 Forward Production systems will
have PDU HC12 5.0.7.
***** - IDC Software 10.0.5 introduced at Penduick must only be used on the Dione 5 form factor IDC. When
changing an IDC on a pre-Penduick IB system, if no more Dione 4 IDCs exist you will receive a Dione 5 IDC on
which you must install IDC SW 10.0.5.