GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
Page no. 1505 Chapter 9
JC-DR-A-321_DS.fm
6-2 Disassembly of IDC Computer on V1/V2 Control Stations
1. Locate the IDC Computer (1) inside the Control Station.
2. Disconnect all cables from the IDC Computer rear panel.
Note:
Note locations of disconnected cables.
3. Disconnect Ground cable (screwdriver).
4. Disconnect the fiber optic cables.
5. For V1 Control Stations, undo the Velcro strips (2) securing the IDC Computer to the Control Station
framework.
1
2

Table of Contents