Art: 730009-00B Rev. Date: 09-Jan-12
2
Note: When running controls, if the handheld is customized for User Defined Reportable
Ranges and Auto Pass/Fail detection, the handheld will display the control result based upon
the customized Reportable Range setting, but will base the Auto Pass/Fail determination on the
measured value, irrespective of the Reportable Range settings.
3.
Narrowing the reportable range of certain tests may affect the presentation of other dependent test
results. In the table below, if any of the results in the first column are outside the reportable range,
the dependent test results listed in the second column will be suppressed (displayed as <>).
Reportable Range Limitations
Test (Outside the Reportable Range) Dependent Suppressed Tests (displayed as <>)
Na K, Cl, BUN, Anion Gap, Hgb, Hct
Hct Cl, BUN, Anion Gap, Hgb
PCO
2
TCO
2
, Anion Gap, Base Excess, HCO
3
, sO
2
pH TCO
2
, Anion Gap, Base Excess, HCO
3
, sO
2
HCO
3
TCO
2
, Anion Gap, Base Excess, sO
2
Cl Anion Gap
K Anion Gap
TCO
2
Anion Gap
PO
2
sO
2
PT/INR PT WBT
MINIMUM SOFTWARE REQUIREMENTS
The minimum handheld and data management requirements for use of the Custom Reportable Range
feature are summarized as follows:
System Component Minimum Software Requirements
i-STAT 1 handheld ≥ JAMS132
Central Data Station Version 5.x ≥ Version 5.26a
i-STAT/DE ≥ Version 2.3