EasyManuals Logo
Home>Abbott>Measuring Instruments>i-STAT 1

Abbott i-STAT 1 System Manual

Abbott i-STAT 1
658 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #204 background imageLoading...
Page #204 background image
14-10 Art: 714376-00M Rev. Date: 31-Jan-12
Control Target Values
and Expected Ranges
Target values (determined by testing multiple vials of each level using
multiple lots of i-STATcartridges with analyzers that have passed the
Electronic Simulator test) are printed on a value assignment sheet
posted on the APOC website at www.abbottpointofcare.com. The ranges
displayed represent the maximum deviation expected when controls
and cartridges are performing properly. Should results outside the range
be obtained, refer to the Troubleshooting portion of this section of the
i-STAT SystemManual.Alwaysbesurethatthelot numberprintedonthe
valueassignmentsheetmatchesthelotnumberonthelabelofthevialinuse,
andthatthesoftwarerevisionabovethetablematchesthesoftwarerevision
intheanalyzer(checkthestatuspageontheanalyzer).
Note: Target values are specific to the i-STAT System; results obtained from these
reconstitutedcontrolplasmasmaydifferifusedwithothermethods.
STEP ACTION
1 After45minuteroomtemperatureequilibration,removethecap
and stopper from one lyophilized human plasma control vial and
remove the cap from one vial of calcium chloride reconstituting
fluid.
2 Pour the entire contents of the calcium chloride vial into the
lyophilized human plasma control vial. Place the stopper back
inthereconstitutedcontrolvial,sealingthevialappropriatelyso
that the contents do not leak or spill out.
3 Allowthevialtositatroomtemperaturefor1minute.
4 Mix the contents of the vial by swirling gently for 1
minute, then inverting slowly for 30 seconds.
Note: Tominimizefoamingofthecontrolsample,avoidvigorous
or rapid mixing motion. Visually inspect the control vial to
ensurethatthesampleisfullyreconstituted.Ifnot,discardthe
reconstituteduidandstartoverwithfreshvials.
5 Usinga plastic transfer pipette, plastic syringe, or plastic
capillarytubewithnoanticoagulant,immediatelytransferthe
solution from the vial into the ACT cartridge
6 Immediately seal the cartridge and insert it into an analyzer.
Note:AdditionalACTcartridgesmaybetestedwiththeremaining
uidifusedwithin30secondsofcompletereconstitutionofthe
sample.
CONTROLS FOR PT/INR CARTRIDGES
Intended Use
The i-STAT
®
PTControlLevel1(normal)andPTControlLevel2(abnormal)
areusedtoverifytheintegrityofnewlyreceivedPT/INRcartridges.
Contents
Each level of control is packaged as a box of 5 vials of lyophilized human
plasma and 5 vials of calcium chloride diluent. The Level 1 control vials contain
9.5
±
1.5mmol/L calcium chloride, whilethe Level 2 controlvials contain
12
±
2.0mmol/Lcalciumchloride.
Storage
i-STAT PT controls, Levels 1 and 2, are contained in 6-mL vials. Separate
6-mL vials contain 1-3 mL of calcium chloride solution for reconstitution.
Refrigerated storage at 2 to 8ºC (35 to 46ºF) should be maintained until the
printed expiration date on the box and vial labels. Do not use beyond the
expirationdateontheboxandviallabels.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals