14-6 Art: 714376-00M Rev. Date: 31-Jan-12
CONTROLS FOR CHEM8+ CARTRIDGES
Control Solutions
Aqueousassayedcontroluidsareavailableforverifyingtheintegrityofnewly
receivedCHEM8+ cartridges.i-STAT CHEM8+Level1, 2and3 Controlsare
formulatedatthreeclinicallyrelevantlevelswithknownconcentrationsof:
Sodium TCO
2
Glucose
Potassium BUN/Urea IonizedCalcium
Chloride Creatinine
Eachlevelofcontrolispackagedinaboxof10ampules.Controlsolutionsare
contained in 1.7 mL glass ampules.
The control solutions do not containhumanserumorserumproducts,butdo
contain buffers and preservatives.
Reactive Ingredients
Storage
Refrigeratedstorageat2to8°C(35to46°F)shouldbemaintaineduntilthe
printedexpirationdateontheboxandampulelabels.
Control solutions may also be stored at room temperature for up to 5 days (18
to30°Cor64to86°F).Prolongedstorageattemperaturesgreaterthan30°C
(86°F)maycausechangesinthevaluesofsomeanalytes.Donotusebeyond
theexpirationdateontheboxandampulelabels.
Best Results
For best results, ampules, cartridges and analyzer should be at the same
temperature.
Ampule Use
A separate ampule must be used for each cartridge being tested.
Donotusethesolutionleftinasyringe,ampuleorcapillarytubeforadditional
testing of cartridges.
Analyte
Calibration
Verification
Level 1
Calibration
Verification
Level 1b
Calibration
Verification
Level 2 and
Control
Level 1
Calibration
Verification
Level 3 and
Control
Level 2
Calibration
Verification
Level 4 and
Control
Level 3
Calibration
Verification
Level 5
Na (mmol/L)
102 102 120 134 160 178
K (mmol/L) 2.20 2.20 2.90 3.90 6.50 8.00
Cl (mmol/L) 70 70 82 95 118 133
Glu (mmol/L) 1.8 1.8 2.5 7.3 17 35
Urea (mmol/L) 45 45 18 4 2.7 1.9
iCa (mmol/L) 2.50 2.50 1.60 1.26 0.80 0.35
Crea (μmol/L)
1477 1477 386 154 46 17
PCO
2
(mmHg)
90 95 66 30 22 18
H
+
(pH)
6.550 6.060 7.005 7.410 7.660 7.920