11-4 Art: 714373-00F Rev. Date: 17-Jan-11
STEP ACTION
Inserting Cartridge into Analyzer
1 Align the cartridge with the contact pads
facing up and toward the cartridge port.
2 Push the cartridge slowly and smoothly into
the cartridge port until it clicks into place.
Removing Cartridge from Analyzer
3 Do not attempt to remove the cartridge
while the message “Cartridge Locked”
remains on the screen.
4 When results are displayed, pull the
cartridge straight out of the analyzer.
5. Dispose of the cartridge in a container
for biohazards, following local, state, and national regulatory
guidelines.
INSERTING AND REMOVING THE CARTRIDGE FROM THE ANALYZER