PROCEDURE MANUAL FOR THE i-STAT SYSTEM
3
REV. DATE: 16-Oct-12 ART: 714446-00L
Cartridges for BNP
Skin puncture: not recommended.
Venipuncture: plastic EDTA collection tubes and disposable transfer device or plastic EDTA syringe.
Blood Volume
See Table 1 below for cartridge volumes.
Table 1: Cartridge Panel Configurations and Blood Volume
(Shading denotes calculated values)
Cartridge
Vol.
L
pH
PCO
2
PO2
Na
K
Cl
iCa
Glu
BUN
Creat
Lact
Hct
TCO
2
ACT
PT/INR
CK-MB
cTnI
BNP
HCO3
TCO2
SO2
BE
Anion Gap
Hb
CHEM8+ 95
CG8+ 95
EG7+ 95
EG6+ 95
CG4+ 95
G3+ 95
EC8+ 65
6+ 65
EC4+ 65
E3+ 65
G 65
Crea 65
ACT 40
PT/INR 20
cTnI 17
CK-MB 17
BNP 17
Suitable Specimens for Cartridges for Blood Gases, Electrolytes, Chemistries, and Hematocrit
Fresh whole blood collected in a plain capillary collection tube or capillary collection tube with balanced heparin.
Fresh whole blood collected in a collection tube with lithium or sodium heparin anticoagulant. Fill collection
tubes to capacity.
Fresh whole blood collected in a plain plastic syringe or in a blood gas syringe labeled for the assays to be
performed. Fill syringes for correct blood-to-heparin ratio.
Suitable Specimens for ACT
Fresh whole blood without anticoagulant collected in a plastic syringe. If from an indwelling line, flush the line
with 5mL saline and discard the first 5mL of blood or six dead space volumes of the catheter.
Fresh whole blood collected in a plastic tube without anticoagulant, clot activators, or serum separators. Device
used to transfer sample to cartridge must be plastic.
Suitable Specimens for PT/INR
Fresh whole blood without anticoagulant collected in a plastic syringe or plastic evacuated tube without clot
activators or serum separators. Device used to transfer sample to cartridge must be plastic.
Fresh capillary whole blood dispensed directly into the cartridge from the finger.
Suitable Specimens for Troponin I/cTnI and CK-MB
Fresh heparinized whole blood or plasma samples collected in syringes or evacuated tubes containing lithium or
sodium heparin. Fill collection tube to capacity.
Non heparinized whole blood samples tested within one minute of patient draw collected into a plastic syringe or
plastic evacuated tube containing no additives.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals