7 - 12
Art: 714369-00H Rev. Date: 03-Aug-12
8. Slide the battery door back onto the compartment until it closes and locks into place.
9. Turn the printer over, plug it back into the AC power adapter, and charge the new battery in the
printer for a minimum of 3 hours before use.
Note: If the rechargeable battery is removed or becomes exhausted, it is still possible to print at reduced
speed using the AC power adapter.
Powering the i-STAT Printer Using the AC Adapter and Power Cord:
1. Connect the power cord to the AC adapter as shown.
2. Plug the round connector from the AC adapter into the 12VDC port on the back of the i-STAT
Printer.
3. Plug the power cord into a wall outlet.
PRINTING DIRECTLY FROM THE i-STAT 1 HANDHELD
1. Ensure that the printer is turned on and that the POWER indicator is green.
2. Align the handheld’s IR communication window with the printer’s IR LED window. Generally, the
printer must be within 1 to 5 inches (2.5 – 12.7 cm) of and not too close to the handheld.
3. Display the results to be printed on the handheld.
4. Press the key on the handheld. Do not move handheld or printer until printing is complete.
5. If printer is not powered from a wall outlet using the AC adapter, turn printer off.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals